Usage
Suitable for radiology department, orthopedics, ward, emergency room, operating room, ICU and other departments.
With flexible movement, easy operation, high efficiency, to meet the radiography for all of human body's parts, such as head, limbs, chest, spine, lumbar vertebrae, abdomen, etc.
Table Movement: X-ray Bed
Parameter | Item |
Digital display SID, tube angle, light touch switch, SCM control, electromagnetic brake | |
Filter | Gobi: 10:1; 103L/INCH; Fixed Filter 15”* 18” |
Height | 650mm |
SID | 500-1280mm |
Tube rotates around the column | -180°~ +180° |
Longitudinal moving of spherical column | 1500mm |
Vertical moving | 930mm |
Lateral movement | 250mm |
Tabletop dimensions | 2200 x 810 mm |
Product Specification
Item | Content | AR-7300A 40kw single phase | AR-7300A 40kw 3 phase | AR-7300C 50kw Single phase | AR-7300C 50kw 3 phase |
Power | Voltage | 220V±22V | 220V±22V 3 phase | 220V±22V | 220V±22V 3 phase |
Frequency | 50Hz±1Hz | 50/60Hz±1Hz | 50Hz±1Hz | 50/60Hz±1Hz | |
Capacity | ≥5kVA | ≥80kVA | ≥5kVA | ≥100kVA | |
Internal Resistance | ≤0.2Ω | ≤0.2Ω | ≤0.2Ω | ≤0.2Ω | |
X-ray Generator | Power | 40KW | 40KW | 50KW | 50KW |
Inverter Frequency | 25KHz | 25KHz | 25KHz | 25KHz | |
Tube voltage | 40kV-125kV | 40kV-125kV | 40kV-150kV | 40kV-150kV | |
Tube current | 10mA-500mA | 10mA-400mA | 10mA-630mA | 10mA-500mA | |
Exposure time | 1.0~6300ms | 1.0~6300ms | 1.0~6300ms | 1.0~6300ms | |
mAs | 0.2mAs-200mAs | 0.2mAs-400mAs | 0.2mAs-200mAs | 0.1mAs-500mAs | |
X-ray tube | Model | TOSHIBA E7239X | TOSHIBA E7239X | XQG50 | XQG50 |
Tube Focus: big/small | 2.0mm /1.0mm | 2.0mm /1.0mm | 1.2mm /0.6mm | 1.2mm /0.6mm | |
Input power | Big focus: 42kW Small focus: 21kW | Big focus: 42kW Small focus: 21kW | Big focus: 50kW Small focus: 20kW | Big focus: 50kW Small focus: 20kW | |
Anode thermal capacity | 100KJ (140KHU) | 100KJ (140KHU) | 220KJ (300Khu) | 220KJ (300Khu) | |
Rotary anode speed | 2700rpm (50Hz) 3200rpm (60Hz) | 2700rpm (50Hz) 3200rpm (60Hz) | 2800rpm | 2800rpm | |
Digital Flat Penal Detector | Active area (HxV mm) | 430×430 | 430×430 | 430×430 | 430×430 |
Pixel matrix (HxV) | 3072×3072 | 3072×3072 | 3072×3072 | 3072×3072 | |
Pixel pitch | 140 μm | 140 μm | 140 μm | 140 μm | |
Cycle time | ≤ 10S | ≤ 10S | ≤ 10S | ≤ 10S | |
Limiting resolution | Min. 3.71 line pair/mm | Min. 3.71 line pair/mm | Min. 3.71 line pair/mm | Min. 3.71 line pair/mm | |
A / D transition | 16 bit | 16 bit | 16 bit | 16 bit | |
Energy range | 40 - 150 kVp | 40 - 150 kVp | 40 - 150 kVp | 40 - 150 kVp | |
Digital Flat Penal Detector – Optional one more. | Active area (HxV mm) | 430 ×430 | 430×430 | 430×430 | 430×430 |
Pixel matrix (HxV) | 3072×3072 | 3072×3072 | 3072×3072 | 3072×3072 | |
Pixel pitch | 139 μm | 139 μm | 139 μm | 139 μm | |
Cycle time | ≤ 10S | ≤ 10S | ≤ 10S | ≤ 10S | |
Limiting resolution | Min. 3.5 line pair/mm | Min. 3.5 line pair/mm | Min. 3.5 line pair/mm | Min. 3.5 line pair/mm | |
A / D transition | 16 bit | 16 bit | 16 bit | 16 bit | |
Energy range | 40 - 150 kVp | 40 - 150 kVp | 40 - 150 kVp | 40 - 150 kVp | |
Work Station | CPU System | Brand: DELL OPTIPLEX 7010 commercial use | |||
Processor: Core™ i5-3570 processor | |||||
RAM: 4GB DDR3 | |||||
Hardware: 500GB SATA (7200 rpm) | |||||
CD-driver: suit MT’s 16X variable-speed DVD + / - RW With functions of double written | |||||
Network Card: Broadcom NetXtreme 10/100/1000 PCIe Gigabit LAN GNTB-A | |||||
High display card: 1GB AMD RADEON HD 7570,FH,with DVI-VGA Adapter | |||||
Slop: PCIE seriel port and parallel port, all height, MT | |||||
Color LCD Monitor | 19 inches 1M LED backing light LCD display | ||||
Revolution1280x1024, 5:4screen | |||||
Dot pitch: 0.294mm | |||||
Contrast ratio: 1000:1 | |||||
Brightness: 250cd/㎡ | |||||
Grey feedback time: 5ms | |||||
Visual Angle: 160/170° | |||||
Work Station Software | Basic operation: Change control console password, edit ID, acquisite images. | ||||
Additional operation: Add new check, edit present checking info, add new position, change image acquisition order, multi-checking agreement chance, manual adjustment of exposure parameters, automatic exposure control mode, focus choice, patient body-type choice, tube capacity check, ESA curve choice, image cutting, note added (sent to DICOM workstation), mark on images, rotate or overturn, full-size image observation, check patient info and dose info, accept or refuse images. | |||||
Image management: Change order, patient basic info editing, inquiry history images, resend history images, re-print history images, check images mark info, review history images, manage refused images, space reclaimed, image protection, and manual image deletion etc. | |||||
System management: ID edition, change ID password, ED refrigeration set, statistics info checking, detector calibration, equipment control, output order management, image measurement. | |||||

Copy product links
Long by picture save/share