Features
AC-5D is the model which can produce average 93% purity oxygen to meet big flow requirements of hospital, home, housecare use and etc. It can stably work in 24 hours a day and needs few maintenances of the main machine among its whole lifetime. Users do not need to open the machine shell to attain filters replacement.
1. 24 hours work continuously
2. Using 2-way oxygen splitter for two peoples
3. Three reliable alarm systems: power off alarm, pressure circulation fault alarm and compressor fault alarm
4. AVG oxygen purity: 93%±3%
5. Low sound level: less than 45dB (A)
6. Net weight: only 15.9kg (for 5 liters)
7. Strong ABS shell: Anti-aging, fashion design
Technical Specifications
PARAMETER | SPECIFICATION |
Oxygen flow | 1-5 LPM Double |
Output pressure | 58.6 kPa |
Power supply | AC220V±22V,50Hz;110V±10V,60Hz |
Input power | 510VA (AVER) |
Oxygen concentration | 1-5 LPM 93%±3% |
Operation Environment | Temperature 5°C-40°C |
Humidity≤80% | |
Atmospheric pressure:86kPa-106 kPa | |
OCSI indicator (optional) | 85% above green |
85% below yellow | |
82% below red | |
Net Weight | 15.9kg |
Noise | <45dB(A) |
Size | 32cm*43cm*62cm(l*w*h) |
Operation regular | Time cycle/pressure rotation |
Transportation/storage condition | Temperature -20°C-55°C |
Humidity≤90% | |
Atmospheric pressure:86kPa-106 kPa | |
Equipment class&Type | Class II B |